单通道移液精准度
移液器 吸头尺寸 量程 系统误差 随机误差 µL %D %CV Flex 1-Channel Pipette (1–50 µL) 20µL
50µL
20µL
50µL
50µL1
1
5
10
50±8.00%
±8.00%
±1.50%
±1.00%
±1.25%3.50%
7.00%
0.75%
0.50%
0.40%Flex 1-Channel Pipette (5–1000 µL) 50µL
50µL
200µL
1,000µL5
5
200
1000±5.00%
±0.50%
±0.50%
±0.50%2.50%
0.30%
0.15%
0.15% -
8通道移液精准度
移液器 吸头尺寸 量程 随机误差 系统误差 µL %D %CV Flex 8-Channel Pipette (1–50 µL) 20µL
50µL
20µL
50µL
50µL1
1
5
10
50±10.00%
±10.00%
±2.50%
±2.50%
±1.25%6.00%
8.00%
1.00%
1.00%
0.60%Flex 8-Channel Pipette (5–1000 µL) 50µL
50µL
200µL
1,000µL5
50
200
1000±8.00%
±2.50%
±1.00%
±0.70%4.00%
0.60%
0.25%
0.15% -
96通道移液精准度
移液器 吸头尺寸 量程 随机误差 系统误差 µL %D %CV Flex 96-Channel Pipette (1–200 µL) 20µL
50µL
20µL
50µL
50µL
200µL1
1
5
5
50
200±10.00%
±10.00%
±4.00%
±4.00%
±1.50%
±1.00%3.00%
6.00%
1.00%
2.00%
1.00%
1.00%Flex 96-Channel Pipette (5–1000 µL) 50µL
50µL
200µL
1,000µL5
50
200
1000±10.00%
±2.50%
±1.50%
±1.50%5.00%
1.25%
1.25%
1.50%